Perform a Control Solution Test:
•BeforetestingwiththeReliOn®microbloodglucosemeter
for the first time.
•Whenyouopenanewbottleofteststrips.
•Wheneveryoususpectthemeterorteststripsmaynotbe
functioning properly.
•Iftestresultsappeartobeabnormallyhighorloworarenot
consistent with clinical symptoms.
•Theteststripbottlehasbeenleftopenorhasbeenexposedto
temperatures below 34°F (1°C) or above 86°F (30°C), or humidity
levels above 80%.
•Tocheckyourtechnique.
•WhentheReliOn®microbloodglucosemeterhasbeen
dropped or stored below 32°F (0°C) or above 122°F (50°C).
Important:
•UseonlywithReliOn®microbloodglucosemeterand
ReliOn® micro blood glucose test strips.